Accident Summary Nr: 14553499 - One employee killed, one injured in lab explosion
| Inspection Nr | Date Opened | SIC | NAICS | Establishment Name |
|---|---|---|---|---|
| 116112194 | 09/18/1995 | 2992 | 0 | J.D. Streett & Co., Inc. |
Abstract: At approximately 2:00 p.m. on September 15, 1995, Employee #1 was installing a new atomic absorption unit. He was not satisfied with the operational flame of the unit and was adjusting the gas flow of acetylene with nitrous oxide when an explosion and fire occurred. Employee #1 was killed. Employee #2, who was performing a lab procedure adjacent to Employee #1, was hospitalized.
| Employee # | Inspection Nr | Age | Sex | Degree of Injury | Nature of Injury | Occupation |
|---|---|---|---|---|---|---|
| 1 | 116112194 | Fatality | Burn/Scald(Heat) | Occupation not reported | ||
| 2 | 116112194 | Hospitalized injury | Burn/Scald(Heat) | Occupation not reported |